| Sample Pre-Treatment | |||
|---|---|---|---|
| Biological Fluids | Mix sample with Lysis-Loading buffer (1:1 or 1:2) | ||
| Tissue | 1. Perform a proteinase K digestion or homogenization of the tissue. 2. Mix sample with Lysis-Loading buffer (1:1 or 1:2) |
||
| SPE protocol | |||
| Solvent / Buffer | Volumes | ||
| Step | 96 Well Plate | 100 mg | 2 mg |
| Condition | Methanol | 1 mL | 200 µL |
| Equilibrate | 50 mM Ammonium Acetate, pH 5.5
*For tissue samples add 0.5% TX-100, 0.1% cysteine to the above-mentioned buffer |
1 mL | 200 µL |
| Load | Pre-treated sample | Up to 0.5 mL | 100 µL |
| Wash | 50 mM Ammonium Acetate, pH 5.5/ Acetonitrile (50:50) | 3x 1mL | 2-3 x 100 µL |
| Elute | 100mM Ammonium Bicarbonate, (pH 9.5) / Acetonitrile / Tetrahydrofuran (50:40:10) | 1mL | 2 x 75 µL |
| Step | Modification |
|---|---|
| Wash | 50mM Ammonium acetate pH 5.5: acetonitrile: MeOH (50:40:10 v/v/v) |
| 50mM Ammonium acetate pH 5.5: acetonitrile (50:50 v/v) | |
| 50mM Ammonium acetate pH 5.5: acetonitrile (30:70 v/v) | |
| 50mM Ammonium acetate pH 5.5: acetonitrile: EtOAC (50:40:10 v/v/v) | |
| Elute | 100mM NH4HCO3 (pH 10): Acetonitrile: THF (50:40:10 v/v/v) |
| 100mM NH4HCO3 (pH 10): Acetonitrile: THF (50:25:25 v/v/v) |