Air bubbles in an HPLC system can disrupt pumping reliability, cause baseline noise, and result in poor result reproducibility and valve function. To prevent these, proper mobile phase degassing is essential. Degassing in HPLC not only ensures stable pumping but also improves sensitivity and accuracy by preventing dissolved oxygen from forming bubbles that interfere with detectors such as UV-VIS, fluorescence, refractive index, or electrochemical. Effective HPLC solvent degassing is therefore a critical step for stable, accurate, and reproducible chromatography.
What is HPLC Solvent Degassing and Why is it Necessary?
When solvents such as water, buffers, acetonitrile, and methanol contact the atmosphere, nitrogen and oxygen dissolve into them. When mixed, their combined air content can exceed the solubility of the mixture, creating a supersaturated condition that leads to outgassing and bubble formation. These bubbles disrupt HPLC performance by causing unstable baselines, bad peak shape, pump cavitation, retention time shifting, and loss of sensitivity.
In pumps, air can stop flowing or cause erratic retention times, while in optical detectors, bubbles scatter light and generate noise or false signals. Although it is not necessary to remove all dissolved air, degassing in HPLC lowers gas levels below supersaturation and prevents these issues. Low-pressure mixing systems are especially prone to bubble formation, whereas high-pressure mixing reduces outgassing because solvents blend under elevated pressure.
Types of Degassing Techniques in HPLC
Degassing the mobile phase was once a major concern in liquid chromatography with outgassing problems being among the most common issues during the first decades of its use. The introduction of inline degassers in HPLC in the past 15–20 years has significantly reduced the occurrence of these problems, to the point that many users today have little experience with them. However, while inline degassing prevents most solvent outgassing issues, it does not eliminate all problems caused by dissolved air.
Inline Degassing
Degassing the mobile phase was once a major concern in liquid chromatography with outgassing problems being among the most common issues during the first decades of its use. The introduction of inline degassers in HPLC in the past 15–20 years has significantly reduced the occurrence of these problems, to the point that many users today have little experience with them. However, while inline degassing prevents most solvent outgassing issues, it does not eliminate all problems caused by dissolved air.
Inline degassing works by passing the mobile phase through a gas-permeable membrane within a vacuum chamber, allowing dissolved gases to escape while retaining the liquid. Different techniques include:
• Vacuum degasser: Inline vacuum degassers, now standard in most modern HPLC systems, have greatly reduced bubble-related pump issues by efficiently removing dissolved gases. Solvents pass through porous tubing within a vacuum chamber, where gases diffuse out through micropores to waste, while the degassed liquid continues flowing onto the pump.
• Membrane degasser: Over the years, various polymer materials have been tried with varying success for degassers. Modern degassers use a thin, highly gas-permeable membrane, making the degassing process very efficient.
Advantages
Inline degassers are highly reliable and effective at reducing dissolved gases in the mobile phase to acceptable levels and are now standard in most HPLC systems. It is usually automatic and trouble-free. They require little to no maintenance, making them convenient and reliable for routine use.
Limitations
Despite their effectiveness, inline degassers do not remove all dissolved air from the mobile phase. The residual air can still cause problems in certain sensitive applications. Membrane replacement, vacuum pump failure, or damaged tubing can also cause issues, so proper maintenance is essential to avoid costly repairs. Additionally, microbial growth in the buffer can block membrane pores and contaminate the entire HPLC system. To prevent microbial growth, contamination, or blockages, it is best not to store the degasser with water or buffer in the tubing and to flush with 30% phosphoric acid followed by water if contamination occurs.