| Possible Cause | Suggested Remedy |
|---|---|
| Detector overload in element specific detectors such as ECD, NPD, FPD, etc. can produce both positive and negative peaks. | Have the compounds of interest arrive at the detector at a different time from the solvent or other compounds in high concentration. H2 produces negative peaks with a TCD and helium carrier gas. |
| Dirty ECD detector can give a negative peak after a positive one. | Clean or replace the ECD detector. |
| Sample contaminants (hydrocarbons or other non-responders) are present when using ECD, PID, or NPD (thermoionic specific) detectors. | Improve sample preparation and cleanup methods prior to injection |
| Possible Cause | Suggested Remedy |
|---|---|
| Incorrect polarity of the recorder connections results in nearly all peaks being negative. | Reverse polarity of recorder connections. |
| Recorder-integrator wires reversed. | Correct connections. |
| Sample injected onto the wrong column for dual-column setups. | Reinject the sample onto the correct column. |
| Possible Cause | Suggested Remedy |
|---|---|
| Detector contamination. | Clean or bake out the detector. |
| Sample contamination. | For PID detectors, check that the sample has not been contaminated with methanol or water. If necessary, prepare a fresh sample. |
| Often normal for NPD (thermoionic specific) detectors. | No correction necessary. |

