How Hot is TOO Hot?

Gas Chromatography Technical Tip

Change Language

French

German

Spanish

Italian

Polish

Czech

For gas chromatographers, dealing with high temperatures is routine. One of the most common questions we have received from customers is: What is the maximum temperature a GC column can reach?
In this technical tip, we continue to cover this topic... along with some recommendations.

Polar stationary phases are more susceptible to damage and degrade at a faster rate when exposed to excessively high temperatures. Thermal damage is evident as excessive column bleed and peak tailing for active compounds. Loss of retention and efficiency (resolution) may be evident for severe cases of thermal damage. Temperature limits do not have any extra safety factors built into the values. Maintaining the column below its upper temperature limits should be practiced to ensure reasonable column life and performance.

Column temperature limits may differ slightly between the same phase with different film thickness. There is an inverse relationship between the film thickness and the upper temperature limit when working with the same stationary-phase. For example:

  • ZB-1 (30m x 0.32mm x 0.25µm) – 360/370 °C iso/grad
  • ZB-1 (30m x 0.32mm x 5.00µm) – 340/360 °C iso/grad
  • But sometimes, even with the phase and the dimensions are the same, the column temperature limits may be different among different manufacturers. Usually, Zebron® columns have limits that are above 10°- 20° competitor columns. Be careful to distinguish between a high temperature version of a stationary phase and the regular version with all the accessories related. A standard silica column may reach up to 360° depending on various parameters as discussed above. The Zebron-HT line is called Inferno™ and you can distinguish them by the different color of the silica tubing that is black and the temperature limit up to 430°.
    If you need to develop a method at higher temperature because you are dealing with very high boiling compounds you have to choose our Zebron-1XT Sim Dist, that has a metal capillary tube and can reach 450° C!

    When working with high temperatures we highly recommend to use only accessories designed for high temperature analysis, like Siltite ferrules, HT Guard Columns and last but not least Zebron Gas Management Filters, which ensure all gases that reach the GC system and the columns, are free from Oxygen and other contaminants that may break the column, especially when working with high temperature and reduce the amount of damage due to the presence of oxygen at high temperature.