| Possible Cause | Suggested Remedy |
|---|---|
| Bad column installation. | Reinstall column. |
| Injector leak. | Find and fix leak. |
| Injection volume too large. | Decrease sample size. |
| Injection temperature too low. | Increase injection temperature so the entire sample is vaporized “instantly”. An injection temperature higher than the temperature limit of the column will not damage the column. |
| Split ratio is too low. | Increase split ratio. |
| Column temperature too low. | Increase column temperature. Use a lower boiling solvent. |
| Initial column temperature too high for splitless injection. | Decrease the initial column temperature. Use a less volatile solvent so the initial column temperature is at least 10°C below the boiling point of the solvent. |
| Purge time too long (splitless injection). | Use a shorter purge activation time. |
| Septum purge plugged or turned off. | Remove plug or turn septum purge on. |