Engineering successful analytical methods using HILIC as an alternative retention mechanism

HPLC / UHPLC Technical Tip

Change Language

French

German

Spanish

Italian

Polish

Czech

Overview: Modern drug candidates present many challenges chromatographically, be them detection or resolution from a co eluting impurity but one of the most common difficulties faced is the retention and resolution of polar compounds that are not amenable to typical C18 reversed-phase conditions.

Many traditional reversed-phase HPLC stationary phases do not provide adequate retention for many of the ever-increasing number of polar compounds analysed by HPLC. These compounds are often either unretained in the void volume of the column, retain so poorly that their peak shape is compromised, or co-elute with other poorly retained peaks. Early-eluting peaks also elute in an ion-suppression zone that is problematic for people using LC-MS instruments.

This tech tip will outline HILIC as an alternative approach that is particularly effective for LC-MS. HILIC uses high levels of organic solvent to elute the polar compounds, which helps to improve their ionization efficiency during LC-MS.

Tip: HILIC works by retaining polar compounds into a water rich layer that is formed at the surface of the silica particle as a result of a polar stationary phase (Fig. 1)

Fig 1.

Acetonitrile (MeCN) is the weak solvent during HILIC and an ammonium buffer the strong solvent. Ammonium salts are critical as a counter ion as it facilitates the formation of the bilayer critical to separation demonstrated in Fig 1.

A polar-rich stationary phase draws a water-rich layer to it’s surface, which promotes retention of polar compounds into the stationary phase. Polar compounds are eluted as you increase the amount of water in the mobile portion of the column, which disrupts the retention of polar analytes along the polar stationary-phase to induce elution.

It is important that you fully equilibrate your HILIC column prior to use to ensure that this water-rich layer has had the appropriate amount of time to form. If this has not occurred, then you will not see consistent retention of your compound or robustness of your method.

We mentioned the use of ammonium buffers, and there are several other factors within a HILIC method that are also critical for its success.

The first is selectivity, for which the correct column is key. As with reversed-phase columns, there are many different stationary phases for HILIC chromatography and they each offer different selectivity profiles and benefits. We would recommend you always endeavour to screen 3 phases of different chemistries to ascertain which offers the best selectivity for your application.

The types of HILIC stationary phases you might consider include:

Regarding the mobile phase, it is important to use MeCN as the organic component – protic solvents such as methanol will disrupt this bilayer formation we spoke about. The amount of organic is also important. Too high of a percentage organic in the initial gradient starting point will complicate and prolong the equilibration of the bilayer; 90% organic is ideal for initial method conditions.

Gradient profile should be 90% MeCN – 50% where possible. This ensures that the bilayer is fully equilibrated at the start of the gradient. If conditions greater than 50% buffer are required to elute the compound, then it is likely the method is not best suited to HILIC, and a polar C18 type of stationary phase might be a better option. You will need to re-equilibrate the bilayer after each gradient run. The higher the buffer percentage at the end of the gradient, the longer this re-equilibration process will take. Incomplete re-equilibration of the buffer will lead to poor reproducibility and a lack of robustness to the method.

The final parameter you should consider carefully is pH. At least 2 different pH’s should be screened, and we would usually suggest pH 3.2 using ammonium formate and 5.8 using ammonium acetate. Different pH’s can lead to significantly different selectivities (Fig. 2).

Fig 2.

Vast changes in elution order between column chemistries and mobile phase conditions

HILIC is an extremely useful tool to improve the retention of polar compounds. The technique can be a challenging approach with which to build robustness, but careful mobile phase choices and method parameters can lead to successful separations for challenging compounds.